Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 59
Brookfi eld Engineering Labs., Inc.
Page 59
Manual No. M08-219-B1211
BREAK: Torque Max
Cause:
The maximum torque has been exceeded during a measurement.
Possible reasons include:
the measuring range is inappropriate for the sample
the bob cannot rotate in the measuring cup
a]
b]
Example
Message
BREAK:torque max!
Single MODE
total MP: 10
total time: 60
What to Do:
press the OK key
check whether the measuring system has been fastened properly at
the rheometer
if fastened properly, select smaller shear rates or speeds for this
measurement or use a measuring system with a higher shear stress
factor
a]
b]
c]
Abort: Speed Max
BREAK: Speed Max
Cause:
The maximum speed has been exceeded during a measurement. Possible
reasons include:
the selected shear stress or torque value for this measurement is too
high
a torque or shear stress measurement has been started without a
sample in the measuring cup
a]
b]
Example
Message
BREAK:speed max!
Single mode
total MP: 10
total time: 60
What to Do:
press the OK key
select smaller shear stress or torque for this measurement
a]
b]