Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 34
Brookfi eld Engineering Labs., Inc.
Page 34
Manual No. M08-219-B1211
If the display fi eld for the torque indicates: “Mlow!”, these values are below the range of resolution
of the rheometer.
If the temperature is not displayed, the measuring sensor is not connected and the temperature
value will show as “1000.0°C” on the printout.
The display is updated with every new measurement point. The current measurement can be
cancelled at any time with the
ST
key.
After a measurement is completed or the
ST
key is pressed, the display fi eld for Step indicates
“END” or “BREAK”. The display alternates at intervals of about four seconds between the last
displayed measurement point and information about the measurement:
Program end
Single mode
Total time: 100s
Total #MP: 10
Reason for
break or end
Duration until
break or end
Type/name of
executed program
Number of measurement
points
Pressing the
OK
key stops the alternating display and returns the user to the menu.
Note: The last selected program parameters remain in the memory even after switching off the
rheometer.
VI.5.3 MAIN
Menu
→ Run Program
The Run Program function starts one of the programs defi ned in memory. If there is no program
available, the following error message is displayed:
No valid program
Enter program fi rst!
In this case, the user has to defi ne a program fi rst (see Section VI.6.2: Utilities
→ Edit Program”).
Use the
OK
key to return to the main menu.
If the programs are available, a list appears to select from. Select the program to be started (see
Section VI.3: Selecting from Lists).
Select PROG to run
1) Prog xyz
2) Test
3) Prog oil 2000