Brookfield R/S Coaxial Cylinder Rheometer User Manual
Page 43
Brookfi eld Engineering Labs., Inc.
Page 43
Manual No. M08-219-B1211
The following values can be edited:
• name of measuring system
• shear rate factor
K [min/s]
(k_gamma)
• shear stress factor ‰ [Pa]
(tau_prom)
Only authorized personnel are permitted to change constants of the
measuring system!
After starting the function, you will be prompted to select the measuring system of which you
want to edit. Select the list item “NEW” to generate a new measuring system.
Select Meas. system:
1) CC40
2) CC25
3) CC14
After selection of the measuring system entry, the user is prompted to enter a name, tau_prom
and k_gamma.
Enter Meas. syst. #4
Name: CC25
tau_prom: 01.1418
k_gamma: 01.2910
The measuring system name is entered in alphanumeric form; the factors tau_ prom and k_gamma
are entered as numerical entries (see Section VI.4).
After entry, the system will prompt the user to store the new parameters.
Press
Name: CC25
tau_prom: 01.1418
k_gamma: 01.2910
Use the OK key to store the new information, otherwise press the ST key to return to the menu
without storing.
VI.6.5 Utilities
→ Print Memory
This function allows the output of the data stored in the instrument memory to either a printer or
to the RS232 serial interface of the rheometer.