Configuration
55
3.2.8
FLDBUS Menu
The FLDBUS menu is shown only if a DeviceNet, Profibus, EtherNet/IP, or ControlNet option card is installed.
The SWAP parameter on the FLDBUS menu enables byte swapping by the iRite BusCommand handler rather than
requiring a SWP (SWAPBYTE) instruction in the PLC. Byte swapping is enabled by default for DeviceNet cards;
for all other fieldbus cards, byte swapping is disabled by default.
•
BYTE swaps bytes within the word before transmission to the scanner
•
WORD swaps words 1 and 2, 3 and 4, within a 4-word packet
•
BOTH performs both operations, swapping bytes within a word and swapping words within the packet
•
NONE disables swapping
The DATASIZE parameter sets the size of the BusCommand handler data transfers. The default value (8 bytes)
matches the default data size specified in the EDS and GSD files, and used by the standard discrete transfer
commands. DATASIZE can be set to any value from 2–128 bytes (1–64 words), but the value specified must match
the data size set for the PLC Scanner I/O data size.
SWAP
NONE
BYTE
DATASIZE
8
data_size
If SWAP = YES
WORD
BOTH
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
VERS
FLDBUS
Figure 3-20. FLDBUS Menu
Table 3-12. Fieldbus Menu Parameters
FLDBUS Menu
Parameter
Choices
Description
Level 2 submenus
SWAP
NONE
BYTE
WORD
BOTH
Specifies byte-swapping used for the fieldbus card. For DeviceNet cards, this parameter defaults to
BYTE; for all other cards the default value is NONE.
NOTE: In Version 3.08 firmware, this parameter supports values of YES (byte swapping) or NO.
Version 3.09 replaces YES with BYTE, NO with NONE, and adds the values WORD and BOTH.
DATASIZE
8
2-128
Specifies the data size, in bytes, that the BusCommand handler transfers. If this parameter is set to
a value other than the default (8 bytes), ensure that it matches the Scanner I/O data size specified
for the PLC.
3.2.9
VERS Menu
The VERS menu can be used to check the installed software version or, by using the
Reset Config
softkey, to restore
all configuration parameters to their factory default values. There are no parameters associated with the Version
menu: when selected, the indicator displays the installed software version number.
Software
version
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
VERS
Figure 3-21. Version Menu
The
Contacts
softkey on the Version menu allows display of contact information (see “FEATURE Menu,
CONTACT Submenu” on page 48). If an
iQUBE
2
scale is configured, a
Diagnostics
softkey also provides access to
iQUBE
2
diagnostic information.