6 enclosure reassembly, 7 cpu board removal, 6 enclosure reassembly 2.7 cpu board removal – Rice Lake 920i USB Installation Manual V5.05 User Manual
Page 22: See section 2.6

16
920i Installation Manual
Expansion Board Serial Port Assignments
Serial port numbers are reserved for each option card
slot, regardless of the type of cards actually installed.
Two port numbers are reserved for each slot that could
contain a dual-channel serial expansion card. Table 2-6
shows the port numbers assigned to each slot.
Table 2-6. Expansion Board Serial Port Assignments
Slot Number
Serial Port Assignments
CPU board
1–4
1
5–6
2
7–8
3
9–10
4
11–12
5
13–14
6
15–16
7
17–18
8
19–20
9
21–22
10
23–24
11
25–26
12
27–28
13
29–30
14
31–32
For example, in a system with a two-card expansion
board, port assignments are reserved as shown in
Figure 2-14. If the only serial card installed in this
system is in SLOT 4 of the expansion board, the system
consists of serial ports 1–4 (on the CPU board) and
ports 11–12.
SLOT 1
SLOT 2
CPU BOARD
1
SLOT 3
SLOT 4
SLOT 0
PORT
1
PORT
2
PORT
3
PORT
4
PORTS
5–6
PORTS
7–8
PORTS
9–10
PORTS
11–12
Figure 2-14. Serial Port Assignments, Two-Card Expansion
Board
2.6
Enclosure Reassembly
Once cabling is complete, position the backplate over
the enclosure and reinstall the backplate screws. Use
the torque pattern shown in Figure 2-15 to prevent
distorting the backplate gasket. Torque screws to 15
in-lb (1.7 N-m).
1
3
5
14
17
16
12
9
8
7
10
11
18
15
4
2
6
13
Torque backplate screws
to 15 in-lb (1.7 N-m)
Figure 2-15. 920i Enclosure Backplate
2.7
CPU Board Removal
If you must remove the 920i CPU board, use the
following procedure:
1. Disconnect power to the indicator. Remove
backplate as described in Section 2.2 on
page 9.
2. Unplug connectors J9, J10, and J11 (serial
communications), J2 (digital I/O), P1 (power
supply), and connectors to any installed option
cards.
3. Remove any installed option cards.
4. Remove the five phillips head screws and two
kep nuts from the CPU board.
5. Gently lift up the CPU board, then disconnect
connectors J12 (power to display), J4 (ribbon
cable, J3 (keypad connector), then the cable J8
(Port 2 serial port).
6. Remove CPU board from the enclosure. If
necessary, cut cable ties to shift cables out of
the way.
To r e p l a c e t h e C P U b o a r d , r e v e r s e t h e a b o v e
procedure. Be sure to reinstall cable ties to secure all
cables inside the indicator enclosure.