6-1
2000-600-02
Manual Loader
The 2000-600-02 Manual Loader is designed as an integral
component of the CDI MULTITEST Torque Calibration System
providing a precision output torque of up to 2000 ft lb while
requiring only 8 ft lb input.
Application
The loader is a laboratory grade instrument used for testing and
calibrating all torque wrenches, drivers, torque multipliers, non-
impact pneumatic and electric nut runners. This equipment also
can be mounted on a fixture for testing and calibrating cable
tensiometers and compression or tension gauges.
Power
To activate the limit solenoids that protect the transducers from
overload, the loader requires AC power, 115/230VAC, 50 to 60
Hz, 3.15 amps.
Do not use this equipment with the power off.
Always turn on the indicator and loader so the
torque and load values are indicated on the display.
The safety relays only work when the power is on.
Transducer Rotation
The manual hand crank provides up to 20° of transducer
rotation in both clockwise (CW) and counter clockwise (CCW)
directions.
MULTITEST Torque Calibration System
Components included in the 2000-600-02 System are:
• 2000-610-02 Indicator,
• 2000-600-02 Loader, and
• 2000-100-02 Roll Cabinet.
When the 2000-600-02 Manual Loader interfaces with the 2000-
610-02 Torque/Force Indicator and the MULTITEST series of
torque transducers, the MULTITEST Torque Calibration system
provides precision loading, monitoring and data acquisition of
torque, force and angle parameters within +/- 0.25% accuracy.
System capabilities include, data storage, retrieval, statistical
analysis and automatic downloading to a printer or computer.