Page 39

background image

3-9

Setup and Programming

DATA LIST Memory

When changing sensors, or when creating a new statistical data

list, the data list memory must be cleared. The entire memory

list or any single memory location may be selectively cleared.

Clearing from the top of the list decrements the total number of

memory locations. Clearing from inside the list shifts all

subsequently taken readings down from that location and

decrements the total number of memory locations.

Clear Last Memory Location

Use the procedure in this section to clear the last memory

location.

1. Press RECALL DATA. The LCD displays as shown in

Figure 3-5. The present memory location is at the upper

left, the total number of memory locations to its right and

the time the data was stored at the far right.

— The LED display shows the measurement value of

that memory location.

— The LCD displays its engineering units at the lower

right.

Figure 3-5: RECALL Data Display

2. Press MANUAL CLEAR. LCD display shows "CLEAR"

and the present (left) and total (right) memory locations

on the upper line, as shown in Figure 3-6.

Figure 3-6: RECALL Data Display

3. Press right DIGIT to select "Single" on the lower LCD

line.