3-4
Setup and Programming
Powering the Equipment
To power the equipment, press the power button on the front of
the MULTITEST Indicator.
• At POWER ON, the LED displays :
CDI
SELFTEST... MEM O.K.
EE O.K.
and the LCD reads:
Firmware version datacode
• With no transducer installed, the LED display reads:
NO INPUT
and the LCD reads:
No transducer
connected
• When applying power with the transducer connected, or when
connecting a transducer while the power is on, the LED
displays:
READ BOX
BOX O.K.
ZERO TARE
+zero
The UNDER, TORQUE and TRACK indicators light. The LCD
displays:
Max=transducer range and units - upper line
x=average and memory location - lower line
4-in-1 Transducer Select
If the 4-in-1 transducer is installed at power on, the LCD display
reads:
Select Transducer
All four transducers are scanned as indicated by their
associated red LED indicators. To select one of the
transducers, apply at least 5% of its full range torque. To return
to the scan mode, press ENTER.
• Then the following banners display:
SOLENOID TEST . . . .
TEST CW
TEST CCW
SELFTEST O.K.