5-12
Calibration
7. Program the measured voltage (example provided in step
5 on the previous page) into the MULTITEST Indicator
using the up and down VALUE keys and the left and right
DIGIT keys.
8. Press ENTER.
— The LED display reads:
APPLY 0V
— LCD display reads:
Calibrate tester - upper line
STEP 2 0V tare - lower line
9. Short pins 5 and 6 together on the calibration fixture.
10.
Using the DIGIT right and left keys and the up and down
VALUE keys enter the following code into the display:
11. Remove the short between pins 5 and 6 of the calibration
fixture.
Press ENTER.
— The LCD display reads
Calibrate Tester - upper line
STEP 3 Adjust mV - lower line
— The LED display reads in millivolts input:
G+XX.XXX
Example:
08.262 mV
12. Using a laboratory millivolt meter, measure the voltage
across pins 5 and 6.
13. Program the measured voltage into the MULTITEST
Indicator as demonstrated in the example, “+08.262” by
pushing and holding the up or down VALUE keys.
14. Press ENTER. The LED display momentarily reads:
SAVING...
15. Turn the power to OFF. Remove the calibration fixture.