Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 58

54
1-4. Protocol
data
All protocol data is ASCII code data.
Transmit format:
<“,”(0x2c)>
data 42><”,”>
”,”(0x2c) delimits each protocol item.
Protocol data download or upload orders:
Item
P.C or Reader Transmit data
1
Protocol mode
“0”(endpoint) or “1”(kinetic), “2”(TeSeE), “3”(TeSeE eq)
2
Security lock status
“0” unlock / “1” lock
3
Kit name[16]:
ASCII characters (max 15characters + null code)
4
Plate reading mode
Plate reading mode (“0”:Single / “1”:Dual)
5
Evaluation mode(endpoint only)
“1”(Normal) or “2”(Evaluation)
6
Precision mode
1:Standard precision mode / 2:High precision mode
7
Number of reads
“1” (endpoint) / “2” to “30”(kinetic mode)
8
Interval time [30](Kinetic mode)
[0]:Start delay time, [1] to [29]:interval time
Start delay: “0” to “999” sec
Interval time: “0” to “1029” sec
9
Measurement wavelength data
“400” to “750”nm
10
Reference wavelength data
“400” to “750”nm
(When the single reading mode is single, Reference wavelength
should be set to “ “(0x20).)
11
Shaking mode
“0” to “2”
“0”(disable)
“1”(1
st
read(kinetic mode only))
“2”(before every read)
12
Shaking speed
“0” to “2”
“0”(Low) / “1”(Middle) / “2”(High)
13
Shaking time
“0”to”999”sec
14
Incubation control temperature
“25.0” to“45.0” (include decimal point)
15
Incubator enabled
“0” or “1”
“0”(No) / “1”(Yes)
16
Cutoff type
“0” to “4”
“0”(Don’t use cutoff)
“1”(cutoff constant)
“2”(cutoff control)
“3”(value by formula)
“4”(Ratio to calibrator)
17
Cutoff value
“0” to “1”
“0”(Single cutoff)
“1”(Ranged cutoff)
18
Cutoff formula type
“1” to “12”
“1”: k*CNx “7”: k + CNx
“2”: k*CPx “8”: k + CPx
“3”: k*COx “9”: k +COx
“4”: CNx / k “10”: k*CNx + CPx
“5”: CPx / k “11”: (CNx+CPx) / k
“6”: COx / k “12”: k1*CNx + k2*CPx
19
Cutoff unit
“0”(STD unit) / “1”(OD)
20
Cutoff K (K and K1)
“0.01” to “10.0”
(data length max: 4 characters(include decimal point))
21
Cutoff K2
“0.01” to “10.0”
(data length max: 4 characters(include decimal point))
22
Cutoff positive const value
“.001 to 9999”