Filter setting, Calendar setting, Lab. name setting – Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 24: Save stdc setting, Save std curve, Rename std curve
20
1:
400
nm 5:650nm
2:450nm 6:...nm
3:475nm 7:...nm
4:650nm 8:...nm
Filter
Day/Month/Year
16
/
05
/
01
Hour: Minutes
18
:
55
Date set
Enter Lab. name
Bio Rad Laboratories
↑
[ABC]
Lab. name
Filter setting
Move the cursor to the desired position and using the Up, Down, and Right
arrow keys. Enter the filter wavelength on the selected position using the
number keys.
If the selected position of the filter disk is empty, enter three “...”.
Press Enter to complete the field entry. The system will return to the previous
Edit Menu screen.
Calendar setting
Move the cursor to the desirable position for date or time input by using the Up,
Down, and Right arrow keys. Enter the present date or time using number
keys.
Press Enter to complete the field entry. The system will return to the previous
Edit Menu screen.
Lab. name setting
Use the Up and Down arrow keys to select from “A” to “Z” and symbol.
Use the Number keys for numeric input.
Right arrow key to move the cursor to the right.
Left arrow key to slip out of the screen and it returns to the Edit Menu screen.
Pressing “./FUNC” key will change the character mode: Upper case alphabet,
Lower case alphabet, and symbol. Current character mode is indicated on the
lower right corner of the LCD.
Save STD Curve
> Save STD Curve
Rename STD Curve
Save STDC
Save STDC setting
Select “Save STDC” in Edit Menu screen above, and press Enter. Then Save
STD Curve Menu shall be appeared on the screen.
Move the cursor to the desired position by using the Up and Down arrow keys.
Select STD storage #
> 1:12/03/04 15:00
2:******** *****
3:15/03/04 10:30
Select for renaming
> 1:16/03/04 11:25
2:****** *****
3:15/03/04 10:30
Save STD Curve
To save a STD Curve, you should select the storage number between #1
and #5. The Up and Down arrow keys are used to move the cursor to the
desired storage number. And press Enter to complete the selection.
After the storage number selection, ”
Are you sure?
” message is appeared.
Select “Yes” or “No” and press Enter. In case of “Yes”, the reader saves
the STD Curve and returns to the Edit Menu screen.
If the curve fit calculation has not done or it has not finished successfully,
an error message will be appeared and saving STD Curve will be
invalidated.
STD curve does not
exist yet.
Press any key
Save STD Curve:
#1:16/03/04 11:25
Are you sure?
> Yes No
Rename STD Curve
To change a name of STD Curve, you should select the storage number
between #1 and #5. The Up and Down arrow keys are used to move the
cursor to the desired storage number. And press Enter to complete the
selection.
Use the Up and Down arrow keys to select from “A” to “Z” and symbol.
Use the Number keys for numeric input.
Right arrow key to move the cursor to the right.
Left arrow key to slip out of the screen and it returns to the Edit Menu
screen.
Pressing “./FUNC” key will change the character mode: Upper case
alphabet , Lower case alphabet , and symbol. Current character mode is
indicated on the lower right corner of the LCD.
#1:16/03/04 11:25
>STD Curve #1 <
[ABC]
Save STD Curve
Rename STD Curve