7.2 Interface
Specifications
The following specification describes the syntax and language required for computer control of the reader.
1. SIGNALS
TXD:
transmit data
RXD:
receive data
DTR:
data terminal ready
DSR:
data set ready
CTS:
clear to send
RTS:
request to send
2. TRANSMISSON
Baud rate:
9600
Data width: 8
Stop width: 1
Parity :
None
3. DATA
All data must be transmitted in ASCII.
The READER command interpreter must be “case blind” (no distinction between upper or lower case
ASCII characters).
The first two characters of the command can distinguish all READER commands from each other.
4. SIGNAL ASSIGNMENT
8
7
6
5
4
3
2
7
8
4
5
6
2
3
Model 680
DSUB 9pin
TXD
RXD
DSR
Signal ground
DTR
CTS
RTS
Host computer
DSUB 9pin
RXD
TXD
DTR
Signal ground
DSR
RTS
CTS
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- iMark Microplate Absorbance Reader (112 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)