Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 53

49
1-2-6 RM (Reset maintenance report)
Reset maintenance report values.
code> Request incubator status from reader. code> status> incubator exist status: “0”:The reader don’t have incubator / “1”:The reader have 1-2-8. INCU (Set Incubator) Incubator ON/OFF control; Benchmark Compatible Format on/off - Incubator ON/OFF code> Set Incubator control temperature data < temperature > Setting temperature value (degree). code> Request filter wheel status from the reader. code> status>
1-2-7. ISTATUS (Get incubator status)
incubator.
on/off - Incubator ON/OFF status, "0": incubator OFF / "1": incubator ON
setting & sensor temperature value:
(Temperature is 3 bytes ASCII encoded decimal value and floating point. ex: "37.0")
"0"; incubator OFF
"1"; incubator ON
1-2-9. WINC (Set Incubator Temperature)
ASCII encoded decimal(1-10)
temperature set range: 25.0 to 45.0 degree
Incubation adjustable in increments of 0.1 degree.
1-2-10. FSTATUS (Get filter wheel status)
wl.1 – wl.8:
filter wavelength at position 1 – position 8. ex.”405”(ASCII encoded decimal)
“***” will be returned when filter is not installed