Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 53

background image

49

1-2-6 RM (Reset maintenance report)

: "RM"

Reset maintenance report values.

: none
: "ERE"

code>


1-2-7. ISTATUS (Get incubator status)

: "ISTATUS"

Request incubator status from reader.

: none
: "ERE"

code>

status>

:

incubator exist status: “0”:The reader don’t have incubator / “1”:The reader have
incubator.
on/off - Incubator ON/OFF status, "0": incubator OFF / "1": incubator ON
setting & sensor temperature value:
(Temperature is 3 bytes ASCII encoded decimal value and floating point. ex: "37.0")

1-2-8. INCU (Set Incubator)

: “INCU

Incubator ON/OFF control; Benchmark Compatible Format

:
:

on/off - Incubator ON/OFF
"0"; incubator OFF
"1"; incubator ON

: "ERE"

code>


1-2-9. WINC (Set Incubator Temperature)

: “WINC

Set Incubator control temperature data

:

< temperature >

:

Setting temperature value (degree).
ASCII encoded decimal(1-10)
temperature set range: 25.0 to 45.0 degree
Incubation adjustable in increments of 0.1 degree.

: "ERE"

code>


1-2-10. FSTATUS (Get filter wheel status)

: "FSTATUS"

Request filter wheel status from the reader.

: none
: "ERE"

code>

status>

:



wl.1 – wl.8:
filter wavelength at position 1 – position 8. ex.”405”(ASCII encoded decimal)
***” will be returned when filter is not installed