4.3.9. Standard setting procedure

Standard curve setting Menu

This menu screen is entered by choosing "STDs" in the End Point Protocol Menu
screen. It is used to set the concentration data and the curve fit type for a standard
curve or to recall a memorized standard curve.
The Up and Down arrow key moves the cursor to the desired item in the menu list. The
Enter key seals the selection and resumes to the next setting screen.


a. STD information setting

STD information

In this screen, there are three items for setting the concentration data of
standards. The Up and Down arrow key moves the cursor to the desired item in
the menu list. The Enter key seals the selection and resumes to the next setting
screen.


Number of Standards

The Numerical keys are used to enter the number of
Standards. 0 means that the curve fit report is not used. The
Enter key sets the number and returns to the previous
screen.
The maximum number of standards is 12.

Concentration

The Numerical keys and the Decimal point key are used to
enter the concentration values of Standards. The Up and
Down
arrow key moves the cursor to the desired position.
The movement of the cursor also accompanies the scroll of
this screen. The Enter key sets all values of the concentration
of standards and returns to the previous screen.

Unit

17 units are supported as below. The unit marked with “[ ]
is the current selection. The Up and Down arrow key moves
the cursor to the desired unit type. The movement of the
cursor also accompanies the scroll of this screen. Press
Enter twice to select the unit type and to return to the
previous screen.

The supported types of unit are:

1. mol/l

11. pg/ml

2. m mol/l

12. U/ml

3. u mol/l

13. IU/ml

4. n mol/l

14. uIU/ml

5. p mol/l

15. mIU/ml

6. mg/l

16. EU

7. ng/l

17. Arbit.

8. pg/l

9. ug/dl

10. ng/ml

STD information

»■Number of STD

•Concentration

•Unit

STD # 1:

0.512

STD # 2:

1.012

STD # 3:

1.512

↓ STD # 4:

2.053

Standard menu

»■STD information
•Curve fit

•Recall STD curve

Concentration

Number of

STD

Number of STD

=

05

(0, 2–12)

»■ 01: ug/dl
02: ng/ml

03:[ pg/ml]

↓ 04: pm/ml

Unit