Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 65

61
Abs. Data Sample of
" 0.101 0.102 0.103 0.104 0.105 0.106 0.107 0.108 0.109 0.110 0.111 0.112"<",">
" 0.201 0.202 0.203 0.204 0.205 0.206 0.207 0.208 0.209 0.210 0.211 0.212"<",">
" 0.301 0.302 0.303 0.304 0.305 0.306 0.307 0.308 0.309 0.310 0.311 0.312"<",">
" 0.401 0.402 0.403 0.404 0.405 0.406 0.407 0.408 0.409 0.410 0.411 0.412"<",">
" 0.501 0.502 0.503 0.504 0.505 0.506 0.507 0.508 0.509 0.510 0.511 0.512"<",">
" 0.601 0.602 0.603 0.604 0.605 0.606 0.607 0.608 0.609 0.610 0.611 0.612"<",">
" 0.701 0.702 0.703 0.704 0.705 0.706 0.707 0.708 0.709 0.710 0.711 0.712"<",">
" 0.801 0.802 0.803 0.804 0.805 0.806 0.807 0.808 0.809 0.810 0.811 0.812"
Example:
eia.reader DRPD ,0,1,End point #1 ,0,415, ,2, ,1,03/2/18 14:00:00,begin, 0.101 0.102 0.103 0.104 0.105
0.106 0.107 0.108 0.109 0.110 0.111 0.112, 0.201 0.202 0.203 0.204 0.205 0.206 0.207 0.208 0.209
0.210 0.211 0.212, 0.301 0.302 0.303 0.304 0.305 0.306 0.307 0.308 0.309 0.310 0.311 0.312, 0.401
0.402 0.403 0.404 0.405 0.406 0.407 0.408 0.409 0.410 0.411 0.412, 0.501 0.502 0.503 0.504 0.505
0.506 0.507 0.508 0.509 0.510 0.511 0.512, 0.601 0.602 0.603 0.604 0.605 0.606 0.607 0.608 0.609
0.610 0.611 0.612, 0.701 0.702 0.703 0.704 0.705 0.706 0.707 0.708 0.709 0.710 0.711 0.712, 0.801
0.802 0.803 0.804 0.805 0.806 0.807 0.808 0.809 0.810 0.811 0.812,end,
1-7-2. Kinetic's Download Raw Plate Data Format
The assignment of the Download Raw Plate Data format of Kinetic:
#
Item of the plate datum
Details
1 Plate data mode
“0”(End point), “1”(Kinetic)
2 Memory number
"1" to "10" (End point), "1" or "2" (Kinetic)
3 Total number of reads
"2" to "30"
4 Reading number
"0", "1" to "30"
(1). The number 0 means that the successive data from here are
the reading condition.
(2). The number 1 to 30 means the reading number of the series of
Kinetic readings.
It should not exceed the total number of reads.
(1) Reading number = "0"
5 Kit name [16]
ASCII characters (max 15characters)
6 Plate reading mode (S/D)
Plate reading mode (“0”: Single / “1”: Dual)
7 Measurement wavelength
Model 680: “400” to “750”nm
8 Reference wavelength
Model 680: “400” to “750”nm
A space character (0x20) is available for Single reading.
9 Measurement filter number
“1” to “8”
10 Reference filter number
“1” to “8”
A space character (0x20) is available for Single reading.
11 Protocol number
"1" to "64" (End point), "65" or "66" (Kinetic)
12 Shaking mode
"0"(disable), "1"(before 1st read), "2"(before every read)
13 Shaking speed
"0"(low), "1"(Middle), "2"(High)
14 Shaking time
"0" to "999" sec
15 Start delay
"0" to "999" sec
16 Reading interval [29]
"6" to "1029" sec for Model 680
Interval time values must be separated by a space code.
1-2,2-3,3-4,4-5,5-6,6-7,7-8,8-9,9-10,10-11,11-12,12-13,13-14,14-1
5,15-16,16-17,17-18,18-19,19-20,20-21,21-22,22-23,23-24,24-25,
25-26,26-27,27-28,28-29,29-30