7.3
Command Language for Model 680 Microplate Reader
7.3.1 Command Language for Remote control mode
1-1. Syntax
SYNTAX:
Command:
"EIA.READER" – Model680 reader
Response:
1-2. Commands
1-2-1. ID (Get Instrument ID)
Request device ID
code> 680 Acquire remote control of the device and lock front panel key pad. code> NOTE: Until the reader receives the “AQ” command, the reader will not accept Release remote control of the device and lock front panel key pad. arguments code> Reset device to power up configuration local mode code> 1-2-5. MR (Maintenance report) Transmit maintenance report code> records; “On/Off:” <#on/off>
1-2-2. AQ (Go to remote control mode)
any other command by responding error code of “8073”-Device not in remote mode.
Once the reader received the “AQ” command, the reader will turn into remote mote
and lock all keys on front panel except for START/STOP for release.
1-2-3. RL (Release remote control mode)
1-2-4. RS (Reset device)
“Hours :” <#hours>
“Plates:” <#plates>
Note: <#on/off>,<#hours> and <#plates> are 4 ASCII encoded decimal values.
This information is kept in battery backup RAM