25

4.3.6. Checkmark value setting (by Administrator only)




















»■Serial #/Report
•Linearity ref.

•Max.CV%

•Filter Check ref.

Checkmark Protocol Menu

»■Serial #:

0000

Rep.type:

Simple

405nm S1:

0.158

Linearity S2:0.309

reference S3:1.401

S4:2.109

Max.CV% »■S1:

15

%

Linearity S2:05%
reference S3:02%

S4:02%

Filter max min

405:

1.875

0.329

450:1.167 0.954

↓ 492:0.596 0.103

Serial #/Report

Linearity reference

Max. CV%

Filter check reference

Filter max min

↑ 540:

1.875

0.329

620:1.167 0.954

690:0.596 0.103

Serial number and Report type setting

Pressing the Up and Down arrow keys will move the cursor and change the
field entry.
When “Serial #” field is active, enter the serial number that will be shown in the
data sheet of the Checkmark plate using the numeral keys.
When the “Rep. type” filed is active, pressing the Right arrow key will switch
the report type between “Simple” and “Full”.

Linearity reference value setting

Pressing the Up or Down arrow keys will move the field entry from S1 to S4.
Pressing the Right arrow key will change the wavelength:
405nm->450nm->492nm->540nm->620nm->690nm
Use the numeral keys for the reference value setting.
Pressing Enter will complete all field entries at all wavelength.

Max. CV% setting

Pressing the Up or Down arrow keys will move the field entry from S1 to S4.
Use the numeral keys for the Max. CV% setting. Pressing Enter will complete
all field entries.

Filter check reference value setting

Initial screen shall be the reference entries for 405nm to 492nm and pressing
the Down arrow key at the bottom of the screen entries will scroll to the next
page for 540nm to 690nm. Pressing the Up, Down and Right arrow keys will
move the field entry.
Use the numeral keys for the filter check reference value setting. Pressing
Enter will complete all field entries.

Checkmark Protocol Menu

Select “Protocol” in the Edit Menu and press Enter. The Checkmark Protocol Menu
will appear on the screen, if the current active protocol type is Checkmark mode.

Edit Menu

»■Protocol •Lab.name

•Security •SaveSTDC

•Filters •Printer
•Date set

Edit Menu

This menu screen can be accessed by pressing the Edit key in the MAIN screen.