4 quick guide – reading a plate – Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 43

39
4.4
Quick Guide – Reading a plate
1. Turn on the instrument. Allow about 15 seconds for the self-diagnosis. Allow the
instrument to warm up for about three minutes prior to reading a plate.
2. Upon power up, the login screen will appear. Enter the login password (initial password:
00000), press Enter.
3. Program a protocol for your assay. See section 4.3.5 for editing a Kinetic protocol, or
section 4.3.7 for editing an End point protocol. Make sure filters installed correlate with
information entered in reader firmware.
4. Select the protocol of choice, if programmed. See the Memory Recall (section 4.3.4) for
detailed operation.
5. Review or print parameters set in the protocol. See section 4.3.7 Editing End Point
Protocol for details.
(a) Verify settings on Cutoff, Report, Limit, STDs, Mode, Mapping, and kit name
parameters are as desired.
(b) Protocol setting can be printed see section 4.3.3 Printing Reports and Protocol
information.
(c) All reports other than the Raw data report requires plate mapping. Set the plate
map with instructions as in section 4.3.11 Plate Map setting procedure.
(d) The Matrix and Limit reports require assignment of upper and lower limits. See
section 4.3.7, Editing End Point Protocol Limit setting window to set the upper
and lower limits.
(e) The Cutoff report requires definition of the cutoff setting. Refer to section 4.3.8
for cutoff setting procedure.
(f) The Curve fit and Concentration reports require that standard concentration and
locations be defined. Refer to section 4.3.9 for details to define the standards
and the curve fit parameters.
(g) Set general reading mode (single or dual wavelength reading, plate shaking,
read mode, and incubation) following instructions in section 4.3.10 Mode setting
procedure.
After warm-up is complete:
6. Press “Main” key, the main screen will appear on the LCD.
7. Carefully place the microplate in the reading chamber. Close the reading chamber door
and press the “Start/Stop” key to read.