4 additional troubleshooting – Bio-Rad QX200™ AutoDG™ Droplet Digital™ PCR System User Manual

Page 42

background image

34 | Automated Droplet Generator

34 | Automated Droplet Generator

Automated Droplet Generator Troubleshooting

4.4 Additional Troubleshooting

Use the following table to help you troubleshoot issues with your AutoDG Instrument. If you have
additional questions or if the recommended action does not resolve the problem, please contact
Bio-Rad Technical Support.

Table 4. Additional troubleshooting for the AutoDG Instrument.
Issue

Possible Causes

Action

No screen display
No response when you

Not connected to power source

Check power cord is securely connected

power on the instrument

Screen display is missing

LCD screen failure

Request service

characters, illegible, or
not responsive to touch

Door won’t open or is

Instrument is running

Wait for droplet generation to complete or touch

hard to open

Instrument is disconnected

the Abort button on the screen to terminate the

from a power supply

run early. Once movement has stopped inside
the instrument, the door will unlock and can be
opened easily.

Connect the instrument to a power supply and wait
for initialization to complete; the door will unlock
and can be opened easily.

Light under a consumable

Consumable is not inserted all

Try positioning the consumable in a different

stays yellow

of the way or is in the incorrect

orientation

orientation

Red Oil Level icon

AutoDG Oil bottle contains

Be prepared to insert a new bottle of oil during the

less than a full plate’s worth of

setup process for the next plate

oil and will require additional oil

to complete the next run