Automated Droplet
Generator Troubleshooting
The Automated Droplet Generator provides onscreen troubleshooting should
an error occur. Guidance will be provided as to how to best resolve the error, or
the instrument will advise that Bio-Rad Technical Support be contacted.
4.1 Restarting a Run after an Error
If the AutoDG
™
Instrument encounters an obstruction during the run, it will
stop the run and display an error message with the area of the deck that
needs attention.
1. Open the door and remove the obstruction and/or replace the consumable
indicated in the onscreen error message.
2. Close the door and touch OK on the error display.
3. Touch the Configure Sample Plate icon on the home screen; the AutoDG
Instrument remembers the status of all consumables so long as it remains
connected to a power supply.
4. On the screen, select the columns of samples remaining to be processed
and touch OK.
5. Touch the Start Droplet Generation icon to resume the run.