beautypg.com

Specifications and maintenance, 1 specifications, Specifications and maintenance 3.1 specifications – Bio-Rad QX200™ AutoDG™ Droplet Digital™ PCR System User Manual

Page 35

background image

Instruction

Manual | 27

26 | Automated Droplet Generator

26 | Automated Droplet Generator

Using the Automated Droplet Generator

Specifications and

Maintenance

3

Weight

100 lb (45.4 kg)

Size (W x D x H)

26 x 22 x 26" (66 x 56 x 66 cm)

Electrical requirements

100–240 V, 50/60 Hz, 90 W; voltage fluctuations not to
exceed +10% of ratings (for external power supply provided)

Fuse

4 A, 24 V internal (not user serviceable)

Temperature

18–30ºC, optimal range is 23 ± 2°C

Altitude

0–6,500 ft (0–2,000 m), optimal range is 0–500 m

Humidity

50% max (noncondensing)

Pollution degree

2 (indoor use)

Installation category

II (external power supply plugs into standard AC receptacle)

Ventilation requirement

5" (13 cm) left and right of machine and 2" (5 cm) behind
should be unobstructed for proper ventilation

3.1 Specifications

26 inches

(66 cm)

35 inches

(89 cm)

open

26 inches

(66 cm)

22 inches

(56 cm)