beautypg.com

2 starting a run on the autodg instrument – Bio-Rad QX200™ AutoDG™ Droplet Digital™ PCR System User Manual

Page 17

background image

Instruction

Manual | 9

8 | Automated Droplet Generator

8 | Automated Droplet Generator

2.2 Starting a Run from Idle Mode on the AutoDG

Using the Automated Droplet Generator

2.2 Starting a Run on the AutoDG Instrument

1. Bring the Automated Droplet Generator out of idle mode by touching the screen. Please wait while the

instrument performs a self-check.

2. Check the indicator lights on the deck of the AutoDG Instrument nd the consumable icons on the touch

screen. Table 3 explains the deck lighting and touch screen icons and the instrument status they indicate.

3. To create a plate of droplets with the Automated Droplet Generator:

a. Touch the Configure Sample Plate button at the bottom center of the screen.

b. Touch or swipe across the screen to select the columns in which your samples are located on the

sample plate. Touching a selected column deselects it. You can touch any orientation of columns.

c. The plate name and plate notes are optional; touching the fields will bring up a keyboard on the screen.

Click OK when done.

The Automated Droplet Generator accepts either a sealed or open 96-well PCR plate
containing 22 µl prepared ddPCR

reactions. Detailed sample preparation and reaction setup

information can be found in the ddPCR Supermix product inserts.

Table 3. AutoDG Instrument status as indicated by deck lighting and touch screen.
Deck Lighting Status

Touch Screen Icon Status

Indication

Off

Gray, Empty

Ready to configure a new run

Off

Gray, Used

Ready to configure a new run; instrument will

prompt for consumable replacement in used
positions when the next run is configured

Green

Green, Ready

Ready to configure a new run; consumables in

the green positions are ready to be used

Yellow

Yellow, Load

Run configured, load consumables as

prompted (this status occurs only during
run setup)

Blue

Blue, Complete

Run complete and droplets ready; occurs

only at droplet plate position