Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 994

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
978
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Selection
Device types | Titrando
Default value
Titrando
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Titrando (without 851, 888, 890), 867
Selection
1 | 2
Default value
1
851, 855, 888, 890
Selection
1
Default value
1
852
Selection
2
Default value
2
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
U(pol)
The polarization voltage is the voltage applied to the polarized electrode
during an amperometric measurement.
Titrando, 855, 867
Range
-1250 ... 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Range
-1270 ... 1270 mV (Increment: 10)
Default value
400 mV
Electrode check
on | off (Default value: off)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.