5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1191
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The filling rate should be reduced for viscous liquids.
5.6.6.3.5.5
DOS U - Monitoring
Tab: Methods
▶ DOS U ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on
this tab:
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this option is switched on then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is trig-
gered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV
Lower hysteresis
Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok is triggered.