7 brc, 1 brc - overview – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 860

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
844
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Selection
Measured value | Time | Water
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV
Time
Range
0.0 ... 999999.9 s
Water
Range
0.0 ... 999999.9 µg
5.6.3.7
BRC
5.6.3.7.1
BRC - Overview
Dialog window: Method
▶ BRC ▶ Properties... ▶ BRC - 'Command name'
Command for coulometric titration for determining the bromine con-
sumption. Out of the amount of bromine consumed the bromine index
(amount of bromine consumed (in mg) per 100 g sample) and the bro-
mine number (amount of bromine (in g) per 100 g sample) can be calcu-
lated.
Instruments
This command can be executed with the following devices.
Titrando: 851, 852
Appearance
The command has the following appearance:
without conditioning
with conditioning
Parameters
The parameters for the command BRC are configured in the following 8
tabs:
■
General/Hardware
Parameters for devices, indicator and generator electrode and stirrers.
■
Start conditions
Entry of a waiting period before the start of the titration.