Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1000

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
984
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.6.6, page 987).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.5 ... 10.0 µA/s
Default value
5.0 µA/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is activated, then the associated time and temperature for
the maximum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.5 ... 10.0 µA/s
Default value
5.0 µA/s
Break point evaluation
Note
This evaluation method is possible only with Titrando, 855 and 867.