5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1133
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Range
0.25 ... 999999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Range
0 ... 4000 mAU
Selection
off
Default value
off
Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:
Duration of measurement
Duration of the adsorption measurement.
Range
0 ... 999999 s
Default value
120 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Range
0.25 ... 999999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Range
0 ... 4000 mAU
Selection
off
Default value
off