5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1081
the drift will be calculated automatically according to the following equa-
tion:
Range
0 ... 999999 s
Default value
52 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.1 ... 999999.0 s
Default value
2.0 s
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Range
-20.0 ... 150°C
Default value
25.0
Switch off stirrer during measurement
Switch off stirrer during measurement
on | off (Default value: off)
If this option is activated, then the stirrer is switched off during the meas-
urement. In this case the following two parameters can be edited:
Stir before measurement
Stirring time between end of dosing and measurement.
Range
0 ... 999999 s
Default value
0 s
Pause before measurement
Waiting time before the measurement during which no stirring takes
place.
Range
0 ... 999999 s
Default value
0 s