6 stat u - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 919

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
903
Selection
off
Default value
off
Titrino
Range
-2000 ... 2000 mV
Selection
off
Default value
off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is not checked until
10 s after the start.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off
5.6.3.8.7.6
STAT U - Stop conditions
Tab: Method
▶ STAT U ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off