6 meas u - fixed endpoint evaluation, 5 meas ipol, 1 meas ipol - overview – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 977: Meas ipol

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
961
[New]
Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 932).
[Properties]
Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 932).
[Delete]
Delete the external measured value selected in the table.
5.6.4.4.6
MEAS U - Fixed endpoint evaluation #
Dialog window: Method
▶ MEAS U ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New]/[Property] ▶ Fixed endpoint evalua-
tion #
Measured quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value (Titrando, 867, 855)
Range
-2000.0 ... 2000.0 mV
Measured value (Titrino)
Range
-2000 ... 2000 mV
Time
Range
0.0 ... 999999.9 s
5.6.4.5
MEAS Ipol
5.6.4.5.1
MEAS Ipol - Overview
Dialog window: Method
▶ MEAS Ipol ▶ Properties... ▶ MEAS Ipol - 'Command
name'
Command for voltametric measurements with selectable polarization
current.