5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
612
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
EP recognition
Filter for the recognition of equivalence points:
Titrino
Selection
all | greatest | last | off
Default value
all
Titrando, 855
Selection
all | greatest | last | ascending | descending |
off
Default value
all
all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.
Evaluation with measured value window (U)
[New]
Open the dialog window Measured value window # in which the
parameters for a new window can be entered (see Chapter 5.6.3.3.3.9,
page 618).
[Properties]
Open the dialog window Measured value window # in which the
parameters for the selected window can be edited (see Chapter
5.6.3.3.3.9, page 618).
[Delete]
Delete the window selected in the table.