3 add - dosing parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1174

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1158
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically
when the dosing has finished. This parameter is displayed only for Titran-
dos, 846, 856, 867, 814, 815, 855, 864 and 874.
5.6.6.2.3
ADD - Dosing parameters
Tab: Method
▶ ADD ▶ Properties... ▶ Dosing parameters
Command name
Name of the command.
Input
25 characters
The volume to be added and the dosing and filling rate are defined on this
tab.
Volume
Fixed volume that is to be dosed.
Titrando, 846, 814, 815, 855, 864, 874
Range
0.00000 ... 99999.9 mL
Default value
10.0000 mL
Titrino
Range
0.0001 ... 9999.0 mL
Default value
10.0000 mL
730, 778, 789
Range
0.001 ... 999.999 mL
Default value
10.0000 mL
774
Range
0.01 ... 999.999 mL
Default value
10.0000 mL