7.20 IO controller
■■■■■■■■■■■■■■■■■■■■■■
1624
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
7.20.8.1.3
General - 2-point controller
The controller acts as a 2-point transfer element.
Behavior of the output
The control deviation represents the difference between setpoint value
and actual value.
7.20.8.1.4
Properties - 2-point controller
Dialogfenster Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'De-
vice type' - 'Device name'
▶ Controller ▶ [Edit] ▶ Properties - 'Controller type'
Controlled variable
Selection of the controlled variable from out of all analog inputs or selec-
tion of a calculated value.
Selection
'Port description' | Calculated value
'Port description'
All analog input ports can be selected.
0% Mapping
Measured value to be used when the Low level (0%) is applied as input
value.
Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.
Range
-214783647.000 ... 214783647.000
Default value
0 or controlled variable setting, respectively
100% Mapping
Measured value to be used when the High level (100%) is applied as
input value.
Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.
Range
-214783647.000 ... 214783647.000
Default value
100 or controlled variable setting, respec-
tively