beautypg.com

Pre-test station checkup, Pre-test station, Checkup – MTS Acumen User Manual

Page 145

background image

Pre-Test Station Checkup

This procedure is recommended anytime you select a new station, change parameter sets or sensor
calibration files with the existing station, or question whether the station is ready to collect valid data. It
describes how to:

Select a controller, configuration, and parameter set.

Verify that the proper calibration file is in use.

Ensure that Station Manager detectors are properly set.

If necessary, create a shortcut that automatically opens the proper controller and the desired Station
Configuration when you launch MPX.

1. Open the Station Manager application.

a) Select All Programs > MTS 793 Software > Station Manager. This launches communication

between the computer and the controller.

MTS Acumen

| 145

Best Practices for Other System Configurations and Tests