beautypg.com

Set limits – MTS Acumen User Manual

Page 107

background image

Set Limits

You can use MPX limit detectors to protect you, your specimen, and the equipment. When a limit is tripped,
an action is carried out (for example, Controlled-Stop Interlock). You can set both the limit values and the
actions.

In this section, you will set nested displacement and force limits. MTS highly recommends mastering the
concepts presented in

Understanding Nested Limits

(p. 80) prior to performing the steps in this topic.

The forces and displacements of interest for this example test are:

Note: This information will be used in the procedure that follows the tables.

Forces of Interest

Maximum Force

Minimum Force

Item

1500 N

-1500 N

Load cell capacity

0 N

-4050 N

Spring capacity

-50 N

-150 N

Tuning procedure forces

0 N

-200 N

Example spring test forces

Displacement Measures of Interest

Displacement Measure

Item

70 mm

Full actuator extension

14 mm

20% actuator extension

56 mm

Remaining actuator extension

90 mm

Full length of beige spring

45 mm

Full compression of beige spring

1. Set tension as positive.

MTS Acumen

| 107

Running the Example Spring Test