2 interface connection – INFICON SKY CDGxxxD (Profibus) User Manual

Page 8

background image

8

tira54e1-a (2009-01) CDGxxxD.cp

2 Interface

Connection

For operating the temperature controlled CDGxxxD gauge via Profibus, an inter-
face cable conforming to the Profibus standard is required.

If no such cable is available, make one according to the following indications.

Only a cable that is suited to Profibus operation may be used (

→ [4] and [5]).

n

Make the Profibus interface cable according to the following indications:

1

5

6

9

D-Sub, 9 pins
male, soldering side

Pin 1 Do not connect

Pin 2 Do not connect

Pin

3 RxD/TxD-P

Pin

4 CNTR-P

1)

Pin

5 DGND

2)

Pin

6 VP

2)

Pin 7 Not connected internally

Pin

8 RxD/TxD-N

Pin 9 Not connected internally

1)

Only to be connected if an optical link module is used.

2)

Only required as line termination for devices at both ends of bus cable

(

→ [4]).

Pin assignment of the D-Sub 15-pins sensor connector according to the
respective operating manual (

→ [1], [2], [3]).

o

Plug the Profibus (and sensor) cable connector into the gauge.

Sensor cable

Profibus cable

p

Lock the Profibus (and sensor) cable connector.

Making a Profibus Interface
Cable

Cable type

Procedure