4 process data – INFICON SKY CDGxxxD (Profibus) User Manual

Page 16

background image

16

tira54e1-a (2009-01) CDGxxxD.cp

The structure of the process data part of the input and output data area depends
on the chosen message telegram. Applicable configurations are described below.

When selecting a message telegram, ascertain what output format of the measured
value (integer/float) is required and whether a parameter channel is needed or not.
The gauge can also be operated in such a way that the master does not transmit
any output data to the slave.

Standard

telegram

Master

Slave

Bytes

Meaning

2

M

⇒ S

0…7 Parameter

Channel

1

S

⇒ M

0

1…2

Exception Status

Process Value UINT16

2

S

⇒ M

0

1…4

Exception Status

Process Value float

7

S

⇒ M

0…7

8

9…10

Parameter Channel

Exception Status

Process Value UINT16

8

S

⇒ M

0…7

8

9…12

Parameter Channel

Exception Status

Process Value float

201

S

⇒ M

0

1…2

3

4…5

Exception Status

Process Value UINT16

Channel Status

Trip Relay 1-2

202

S

⇒ M

0

1…4

5

6…7

Exception Status

Process Value float

Channel Status

Trip Relay 1-2

203

S

⇒ M

0…7

8

9…10

11

12…13

Parameter Channel

Exception Status

Process Value UINT16

Channel Status

Trip Relay 1-2

204

S

⇒ M

0…7

8

9…12

13

14…15

Parameter Channel

Exception Status

Process Value float

Channel Status

Trip Relay 1-2


4.4 Process Data

4.4.1 Standard Telegrams