INFICON SKY CDGxxxD (Profibus) User Manual
Page 10

10
tira54e1-a (2009-01) CDGxxxD.cp
THE "NET" LED indicates the network status of the CDGxxxD gauge:
LED
Status
Off
Device is off
Red
Error. The alarm bit in Exception Status is
set. See section 7.1.11.
Green/flashing
Device is ok and uses acyclic data traffic.
See section 5.1.
Green
Device is ok and uses cyclic data traffic.
See section 5.2.
The gauge supports all data rates defined in the IEC 61158 Type 3 / IEC 61784
standards (
→ [5]) up to 12 Mbaud. Automatic data rate setting is supported.
Alternatively, a fixed data rate can be selected.
The ident number assigned to the gauge by the PNO (
→ [4]) is:
Gauge
Ident number (hexadecimal)
CDGxxxD 0B7E
Depending on the standard telegrams used (
→ section "Cyclic Message Tele-
grams"), the following configuration data have to be transmitted to the gauge during
the configuration phase:
Standard
telegram
M
⇒
S
Standard
telegram
S
⇒
M
M
⇒
S
S
⇒
M
Configuration data
ParamChannel
ParamChannel
Exception Status
Process Value UINT
16
Process Value float
Channel Status
Trip Relay 1-2
1
x
x
0x42,0x82,0x05,0x03
2
x x
0x42,0x84,0x05,0x08
2
7
x x x x 0xC4,0x87,0x8A,0x0A,0x0A,0x05,0x03
2
8
x x x x
0xC4,0x87,0x8C,0x0A,0x0A,0x05,0x08
201
x x x x
0x44,0x85,0x05,0x03,0x05,0x0A
202
x x x x
0x44,0x87,0x05,0x08,0x05,0x0A
2
203
x x x x x x
0xC6,0x87,0x8D,0x0A,0x0A,0x05,0x03,0x05,0x0A
2
204
x x x x x x
0xC6,0x87,0x8D,0x0A,0x0A,0x05,0x08,0x05,0x0A
3.4 "NET" LED
3.5 Data Rate
3.6 Ident Number
3.7 Configuration Data