2 atm transducer block (instance 2) – INFICON SKY CDGxxxD (Profibus) User Manual
Page 44

44
tira54e1-a (2009-01) CDGxxxD.cp
ID Name
Structure Data type
Bytes Access
Store See section
14 Atmosphere
Sensor
Block
Command
Array UINT8
2 1_R/W
2_R/W
N 12.2.5
101 Block
Type
ID
Simple OSTRING(n) 4 1_R
2_R
N 12.2.1
102 Status
Extension
Simple UINT8
1 1_R
2_R
V 12.2.2
103 Sensor
Alarm
Array UINT8
2 1_R
2_R
V 12.2.3
104 Sensor
Warning
Array UINT8
2 1_R
2_R
V 12.2.4
The Block Type ID Parameter contains an ID which describes the block type. The
block type ID of the "ATM Transducer Block" is set to "256". The other defined
block types are listed in Appendix B.
This bit mapped byte parameter provides the following additional information of the
ATM sensor (Instance 2):
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Res. Res.
Res.
Res.
Res.
Underrange
Exceeded
Overrange
Exceeded
Reading
Invalid
Res. = Reserved
• Underrange
Exceeded
This bit is set ("1") if the ProcessValue PV is below the parameter Underrange
(ID 45) of the ATM sensor instance. See section 8.2.11.
• Overrange
Exceeded
This bit is set ("1") if the ProcessValue PV is above the parameter Overrange
(ID 44) of the ATM sensor instance. See section 8.2.10.
• Reading
invalid
This bit represents the logical inversion of the parameter "Reading Valid"
(ID 23) of the ATM sensor instance. See section 8.2.6.
This byte is always "0". (The ATM Sensor does not produce any errors).
This byte is always "0". (The ATM Sensor does not produce any warnings).
12.2 ATM Transducer Block
(Instance 2)
12.2.1 Block Type ID (ID 101)
12.2.2 StatusExtension
(ID 102)
12.2.3 Sensor Alarm (ID 103)
12.2.4 Sensor Warning (ID 104)