beautypg.com

8 sensor analog input function blocks – INFICON SKY CDGxxxD (Profibus) User Manual

Page 27

background image

tira54e1-a (2009-01) CDGxxxD.cp

27

8 Sensor Analog Input Function Blocks

All gauge functions of the CDGxxxD are described in the Analog Input Function
Block. Because the gauge includes two measuring systems, there are two Analog
Input Function Block Instances:
• Instance 1 represents the Capacitance Diaphragm Gauge (CDG) measuring

part of the gauge. See section 8.1.

• Instance 2 represents the Atmosphere Pressure Sensor (ATM) measuring part

of the gauge. See section 8.2.

Instance 1 of the Analog Sensor Input Function Blocks describes the functionality
of the Capacitance Diaphragm Gauge (CDG) measuring part of the gauge. The
following attributes are supported:

ID Name

Structure Data type

Bytes Access Store See section

15 AIBlockAdjust

Command

Record UINT8

2 1_R/W

2_R/W

N 8.1.22

16 Block Type ID

Simple

OSTRING(n)

4

1_R
2_R

N 8.1.1

17 AlarmEnable

Simple Boolean

1 1_R/W

2_R/W

N 8.1.2

18 WarningEnable

Simple Boolean

1 1_R/W

2_R/W

N 8.1.3

19 ProcessValue

(PV)

Simple *)

- 0_XI

1_R
2_R

D 8.1.4

20 Status

Simple UINT8

1 0_XI

1_R
2_R

D 8.1.5

21 Data

Type

Simple UINT8

1 2_R/W N 8.1.6

22 Data

Unit

Simple UINT16

2 2_R/W N 8.1.7

23 Reading

Valid

Simple Boolean

1 1_R

2_R

D 8.1.8

24 Full

Scale

Simple **)

- 1_R

2_R

N 8.1.9

25 Offset-A Data Type

Simple

UINT8

1

1_R
2_R

N 8.1.10

26 Offset-A

Simple **)

- 1_R

2_R

8.1.11

31 Alarm Trip Point High

Simple

**)

-

1_R/W
2_R/W

N 8.1.12

32 Alarm Trip Point Low

Simple

**)

-

1_R/W
2_R/W

N 8.1.13

33 Alarm

Hysteresis

Simple **)

- 1_R/W

2_R/W

N 8.1.14

35 Warning Trip Point High

Simple

**)

-

1_R/W
2_R/W

N 8.1.15

36 Warning Trip Point Low

Simple

**)

-

1_R/W
2_R/W

N 8.1.16

37 Warning

Hysteresis

Simple **)

- 1_R/W

2_R/W

N 8.1.17

39 Safe

State

Simple **)

1 1_R/W

2_R/W

N 8.1.18

40 Safe

Value

Simple **)

- 1_R/W

2_R/W

N 8.1.19

44 Overrange

Simple **)

- 1_R

2_R

N 8.1.20

45 Underrange

Simple **)

- 1_R

2_R

N 8.1.21

8.1 Sensor Analog Input

Function Block
(Instance 1, CDG)