8 Sensor Analog Input Function Blocks
All gauge functions of the CDGxxxD are described in the Analog Input Function
Block. Because the gauge includes two measuring systems, there are two Analog
Input Function Block Instances:
• Instance 1 represents the Capacitance Diaphragm Gauge (CDG) measuring
part of the gauge. See section 8.1.
• Instance 2 represents the Atmosphere Pressure Sensor (ATM) measuring part
of the gauge. See section 8.2.
Instance 1 of the Analog Sensor Input Function Blocks describes the functionality
of the Capacitance Diaphragm Gauge (CDG) measuring part of the gauge. The
following attributes are supported:
ID Name
Structure Data type
Bytes Access Store See section
15 AIBlockAdjust
Command
Record UINT8
2 1_R/W
2_R/W
N 8.1.22
16 Block Type ID
Simple
OSTRING(n)
4
1_R
2_R
N 8.1.1
17 AlarmEnable
Simple Boolean
1 1_R/W
2_R/W
N 8.1.2
18 WarningEnable
Simple Boolean
1 1_R/W
2_R/W
N 8.1.3
19 ProcessValue
(PV)
Simple *)
- 0_XI
1_R
2_R
D 8.1.4
20 Status
Simple UINT8
1 0_XI
1_R
2_R
D 8.1.5
21 Data
Type
Simple UINT8
1 2_R/W N 8.1.6
22 Data
Unit
Simple UINT16
2 2_R/W N 8.1.7
23 Reading
Valid
Simple Boolean
1 1_R
2_R
D 8.1.8
24 Full
Scale
Simple **)
- 1_R
2_R
N 8.1.9
25 Offset-A Data Type
Simple
UINT8
1
1_R
2_R
N 8.1.10
26 Offset-A
Simple **)
- 1_R
2_R
8.1.11
31 Alarm Trip Point High
Simple
**)
-
1_R/W
2_R/W
N 8.1.12
32 Alarm Trip Point Low
Simple
**)
-
1_R/W
2_R/W
N 8.1.13
33 Alarm
Hysteresis
Simple **)
- 1_R/W
2_R/W
N 8.1.14
35 Warning Trip Point High
Simple
**)
-
1_R/W
2_R/W
N 8.1.15
36 Warning Trip Point Low
Simple
**)
-
1_R/W
2_R/W
N 8.1.16
37 Warning
Hysteresis
Simple **)
- 1_R/W
2_R/W
N 8.1.17
39 Safe
State
Simple **)
1 1_R/W
2_R/W
N 8.1.18
40 Safe
Value
Simple **)
- 1_R/W
2_R/W
N 8.1.19
44 Overrange
Simple **)
- 1_R
2_R
N 8.1.20
45 Underrange
Simple **)
- 1_R
2_R
N 8.1.21
8.1 Sensor Analog Input
Function Block
(Instance 1, CDG)