INFICON SKY CDGxxxD (Profibus) User Manual
Page 35

tira54e1-a (2009-01) CDGxxxD.cp
35
This parameter contains the lowest valid Process Value (PV) of the device in terms
of the currently selected data type (ID 21) and data unit (ID 22).
Sensor Type
COUNTS
Torr
mbar
Pascal
1000 Torr
3
100
1.333224 × 10
-1
13.33224
500 Torr
3
50
6.66612 × 10
-2
6.66612
200 Torr
3
20
2.666448 × 10
-2
2.666448
100 Torr
3
10
1.333224 × 10
-2
1.333224
50 Torr
3
5
6.66612 Ч 10
-3
6.66612 Ч 10
-1
20 Torr
3
2
2.666448 Ч 10
-3
2.666448 Ч 10
-1
10 Torr
3
1
1.333224 Ч 10
-3
1.333224 Ч 10
-1
5 Torr
3
0.5
6.66612 Ч 10
-4
6.66612 Ч 10
-2
2 Torr
3
0.2
2.666448 Ч 10
-4
2.666448 Ч 10
-2
1 Torr
3
0.1
1.333224 Ч 10
-4
1.333224 Ч 10
-2
500 mTorr
3
0.05
6.66612 Ч 10
-5
6.66612 Ч 10
-3
200 mTorr
3
0.02
2.666448 Ч 10
-5
2.666448 Ч 10
-3
100 mTorr
3
0.01
1.333224 Ч 10
-5
1.333224 Ч 10
-3
1000 mbar
3
7.500615 Ч 10
-2
1 Ч 10
-1
10
500 mbar
3
3.750308 Ч 10
-2
5 Ч 10
-2
5
200 mbar
3
1.500123 Ч 10
-2
2 Ч 10
-2
2
100 mbar
3
7.500615 Ч 10
-3
1 Ч 10
-2
1
50 mbar
3
3.750308 Ч 10
-3
5 Ч 10
-3
5 Ч 10
-1
20 mbar
3
1.500123 Ч 10
-3
2 Ч 10
-3
2 Ч 10
-1
10 mbar
3
7.500615 Ч 10
-4
1 Ч 10
-3
1 Ч 10
-1
5 mbar
3
3.750308 Ч 10
-4
5 Ч 10
-4
5 Ч 10
-2
2 mbar
3
1.500123 Ч 10
-4
2 Ч 10
-4
2 Ч 10
-2
1 mbar
3
7.500615 Ч 10
-5
1 Ч 10
-4
1 Ч 10
-2
0.5 mbar
3
3.750308 Ч 10
-5
5 Ч 10
-5
5 Ч 10
-3
0.2 mbar
3
1.500123 Ч 10
-5
2 Ч 10
-5
2 Ч 10
-3
0.1 mbar
3
7.500615 Ч 10
-6
1 Ч 10
-5
1 Ч 10
-3
ID
Name
Structure
Data Type
Bytes
Access
Store
15 AIBlockAdjust
Command
Record UINT8 2 1_R/W
2_R/W
N
Byte
Name
Structure
Data Type
Bytes
Access
Store
0 Adjust
Command
Simple UINT8 1 1_R/W
2_R/W
N
1 Target
Value Simple
*)
1 1_R/W
2_R/W
N
*) According to data type value (ID 21)
Adjust Command
Name
Description
0…255
Reserved
Reserved, no action
Target value
Description
0…255
Reserved, no action
8.2.11 Underrange (ID 45)
8.2.12 Analog Input Block
Adjust Command
(ID 15)