Chapter 3 menus – INFICON SQC-310 Thin Film Deposition Controller User Manual
Page 61
Chapter 3
Menus
3-23
Pocket Wait: This parameter has two functions, depending on the setting of the Pocket
Ready signal in I/O setup. If no Pocket Ready input is programmed for an output, then
this is the time the program waits for the indexer to complete its move. After the wait
time, deposit begins.
If a Pocket Ready input is programmed for an output, then this is a timeout value. If the
pocket ready signal is not received within Time Wait, then an error occurs.
Dev Graph Limit: Sets the upper limit for the Rate Deviation graph Y axis.
Rate Alarm Low: Sets the % deviation below rate setpoint that causes an audible
alarm.
Rate Alarm High: Sets the % deviation above rate setpoint that causes an audible
alarm. The tone of the high rate alarm is lower than that of the low rate alarm.
Password Enable: If Password is enabled, the Quick Start, Film and System Menus
require a password. The Process Menu can be used to select a process, but a
password is required to make any changes on the Process Menu.
Password: If password is enabled, this parameter sets the sequence of SoftKeys to
press to enter menus. Press the desired sequence to set the password.
Note: Holding down the top and bottom switch while powering up the SQC-300 sets the
password to 1111.
System Time: Sets the SQC-300 clock to your local time. System time is battery
backed up. Does not compensate for daylight savings time.
System Date: Sets the SQC-300 date to your local time. System date is battery
backed up.