Maintenance steps 7.4.1, 4 maintenance steps – INFICON LDS3000 Modular Leak Detector User Manual

Page 45

background image

Maintenance

45

7.4

Maintenance steps

7.4.1

Change operating fluid reservoir of turbo molecular
pump

1

Shut down mass spectrometer module, see

Chapter 8, page 49

.

2

Remove turbo molecular pump.

Flooding mass spectrometer and turbo molecular pump

The cover of the operating fluid reservoir cannot be unscrewed unless the reservoir
has been flooded.

1

Disconnect 24 V power supply pack from MSB box.

2

Wait until turbo molecular pump is drained (at least 1 min).

3

Allow the turbo molecular pump to cool down if necessary.

4

Open the ventilation screw slowly.

 Turbo molecular pump is flooded until it reaches atmospheric pressure.

Fig: 5

Turbo molecular pump SplitFlow 80

1

Cover

2

O-ring

3

Operating fluid reservoir

4

Ventilation screw