Activate/deactivate zero, Selecting a signal filter, 8 activate/deactivate z – INFICON LDS3000 Modular Leak Detector User Manual
Page 31: 9 selecting a signal filter
Operation
31
6.8
Activate/deactivate Z
ERO
Z
ERO
can be used to suppress undesired helium backgrounds. If Z
ERO
is enabled,
the currently measured leakage rate value will be interpreted as a helium background
and subtracted from all subsequently measured values.
The background value suppressed by Z
ERO
is adjusted automatically if the
background changes inside the unit.
Information If the signal filter Fixed or 2-Zone is selected, then the background value
is automatically adjusted based on the configured zero time.
► Enable or disable Z
ERO
as needed:
S
ETTINGS
> Z
ERO
/F
ILTER
> Z
ERO
> Z
ERO
MODE
Z
ERO
WITH
S
TART
suppresses the helium background automatically when a
measurement is started.
► Enable or disable Z
ERO
WITH
S
TART
as needed:
S
ETTINGS
> Z
ERO
/F
ILTER
> Z
ERO
> Z
ERO
WITH
S
TART
6.9
Selecting a signal filter
► As a rule, use signal filter I•CAL.
► If the signal filter should simulate the time behavior of older units, then use filter
Fixed or 2-Zone.
► Select desired signal filter:
S
ETTINGS
> Z
ERO
/F
ILTER
> F
ILTER
> F
ILTER
MODE
Signal filter
I•CAL
The leakage rates are averaged at time intervals that are optimized
for the range of the leakage rates.
Fixed
The leak rates are averaged at fixed intervals of 0.2 seconds.
2-Zone
Filter compatible with LDS1000 and LDS2000
The averaging period is switched depending on the leak rate
threshold.