Installation

19

5.3

Connecting the mass spectrometer module to the test system

The operation mode of the vacuum connection and the speed of the turbo molecular
pump define:

• Minimum detectable leak rate

(MDLR)

• Maximum inlet pressure

(P

max

)

• Volume flow rate

(S)

* constantly permissible inlet pressure

Temporary exceed (< 3 s):

Permissible inlet pressure is 0.4 mbar in U

LTRA

and 0.7 mbar in F

INE

.

Long-term exceed (> 3 s):
Warning (TMP overheating) and possible damage to the filament.

1

Set the operation mode vacuum connection and the speed turbo molecular pump
in accordance with the physical vacuum conditions found in the test system.

2

Connect the mass spectrometer module to the U

LTRA

, F

INE

or G

ROSS

connections

on the vacuum system of the test system.

3

Set the speed of the turbo molecular pump.

Connection

Speed turbo molecular pump

1000 Hz

1500 Hz

U

LTRA

MDLR: < 5 x 10

-12

mbar·l/s

< 1 x 10

-11

mbar·l/s

P

max

: 0.2 mbar

0.2 mbar *

S: 5 l/s

6 l/s

F

INE

MDLR: < 1 x 10

-11

mbar·l/s

< 5 x 10

-11

mbar·l/s

P

max

: 0.9 mbar

0.4 mbar *

S: 1.8 l/s

2.5 l/s

G

ROSS

MDLR: < 1 x 10

-9

mbar·l/s

< 2 x 10

-8

mbar·l/s

P

max

: 18 mbar

15 mbar

S: depends on the backing pump

Material damage due to pressure surges

Pressure surges exceeding the maximum inlet pressure will damage the mass
spectrometer module.
► Do not exceed the maximum inlet pressure.