5.3
Connecting the mass spectrometer module to the test system
The operation mode of the vacuum connection and the speed of the turbo molecular
pump define:
• Minimum detectable leak rate
(MDLR)
• Maximum inlet pressure
(P
max
)
• Volume flow rate
(S)
* constantly permissible inlet pressure
Temporary exceed (< 3 s):
Permissible inlet pressure is 0.4 mbar in U
LTRA
and 0.7 mbar in F
INE
.
Long-term exceed (> 3 s):
Warning (TMP overheating) and possible damage to the filament.
1
Set the operation mode vacuum connection and the speed turbo molecular pump
in accordance with the physical vacuum conditions found in the test system.
2
Connect the mass spectrometer module to the U
LTRA
, F
INE
or G
ROSS
connections
on the vacuum system of the test system.
3
Set the speed of the turbo molecular pump.
Connection
Speed turbo molecular pump
1000 Hz
1500 Hz
U
LTRA
MDLR: < 5 x 10
-12
mbar·l/s
< 1 x 10
-11
mbar·l/s
P
max
: 0.2 mbar
0.2 mbar *
S: 5 l/s
6 l/s
F
INE
MDLR: < 1 x 10
-11
mbar·l/s
< 5 x 10
-11
mbar·l/s
P
max
: 0.9 mbar
0.4 mbar *
S: 1.8 l/s
2.5 l/s
G
ROSS
MDLR: < 1 x 10
-9
mbar·l/s
< 2 x 10
-8
mbar·l/s
P
max
: 18 mbar
15 mbar
S: depends on the backing pump
Material damage due to pressure surges
Pressure surges exceeding the maximum inlet pressure will damage the mass
spectrometer module.
► Do not exceed the maximum inlet pressure.