Assigning the digital outputs of the i/o module, 4 assigning the digital outputs of the i/o module – INFICON LDS3000 Modular Leak Detector User Manual
Page 28
28
Operation
Information Every function can be inverted.
► Assign the digital inputs of the I/O module as needed:
S
ETTINGS
> S
ET
UP
> I
NTERFACES
> I/O
MODULE
> PLC
INPUTS
>
C
ONFIGURATION
DIG
. I
NPUT
6.4
Assigning the digital outputs of the I/O module
The available functions can be assigned in any way necessary to the digital outputs
PLC-OUT 1 ... 8 of the I/O module:
STOP
inactive→ active:
Switch to Standby.
(Z
ERO
is not possible, all trigger outputs will return "Fail".)
ZERO
inactive→ active:
active→ inactive:
Switch Z
ERO
on.
Switch Z
ERO
off.
ZERO_PULS
inactive→ active. Switch Z
ERO
on or off.
CLEAR
inactive→ active:
Erase warning or error message / cancel calibration.
GASBALLAST
inactive→ active:
active→ inactive:
Open gas ballast valve.
Close gas ballast valve unless always open.
SELECT_DYN_NORMAL
inactive→ active:
active→ inactive:
External calibration mode for the "CAL" function:
External dynamic calibration
(without auto tune, allowing for the measuring times and pump
cycle times set via the digital inputs)
External normal calibration
(with auto tune, not considering the system-specific measuring
times and pump cycle times)
START_STOP
inactive→ active:
active→ inactive:
Switch to Meas.
(Z
ERO
is possible, all trigger outputs switch depending on the
leakage rate.)
Switch to Standby.
(Z
ERO
is not possible, all trigger outputs will return "Fail".)
KEY_1
active:
User "Operator"
KEY_2
active:
User "Supervisor"
KEY_3
active:
User "Integrator"
CAL
inactive→ active:
When set to Standby, the unit will start an internal calibration.
When set to Meas, the unit will start an external calibration.
Function
Flank/state:
Description
Function
State:
Description
OPEN
open:
always open
TRIG1
closed:
open:
Value exceeded leak rate threshold Trigger 1
Value fell below leak rate threshold Trigger 1
TRIG2
closed:
open:
Value exceeded leak rate threshold Trigger 2
Value fell below leak rate threshold Trigger 2
TRIG3
closed:
open:
Value exceeded leak rate threshold Trigger 3
Value fell below leak rate threshold Trigger 3
TRIG4
closed:
open:
Value exceeded leak rate threshold Trigger 4
Value fell below leak rate threshold Trigger 4
READY
closed:
open:
Emission switched on, calibration process inactive, no error
Emission switched off, calibration process active or error