Operation
41
* compatib. LDS1000/LDS2010 or binary protocol
611
W
Machine factor too high,
(machine factor)
81
153
1.00E+
04
Machine factor adjustment
inaccurate, split flow factor
too high
612
W
Machine factor lower
than last,
(machine factor)
81
153
< 50%
old
value
Split flow factor has
changed
613
W
Machine factor higher
than last,
(machine factor)
81
153
> 200%
old
value
Split flow factor has
changed
625
W
Int testleak not set,
(int. Testleak)
0
0
Leakage rate of internal
test leak is still set to
factory setting
626
W
Ext testleak not set,
(ext. Testleak)
0
0
Leakage rate of test leak is
still set to factory setting
630
W
Calibration request
0
0
Temperature change of
5°C, speed was changed
since last calibration, 30-
minute switch-on time and
still no calibration
conducted
7xx temperature errors (pre-amplifier, electronics)
700
W
Preamplifier temp too
low,
(temperature [°C])
33
60
2°C
Temperature too low
702
W
Preamplifier temp too
high,
(temperature [°C])
32
124
60°C
Temperature too high
703
E
Preamplifier not
connected
33
60
Pre-amplifier defective,
cable not plugged in
710
W
Max. MSB temperature
exceeded,
(temperature [°C])
54
44
55°C
Temperature too high
711
E
MSB temperature far
too high,
(temperature [°C])
54
44
65°C
Temperature too high
8xx not used
9xx Maintenance warnings (e. g. TMP)
901
W
Maintenance bearing/
lubricant
99
99
3 years
TMP maintenance
necessary
Error
number
Warning/
Error
Error message
LDS3000
Error
number
LDS1000
protocol
Error
number
ASCII
protocol*
Limit
values
Cause