Operation

29

Information Every function can be inverted.

► Assign the digital outputs of the I/O module as needed:

S

ETTINGS

> S

ET

UP

> I

NTERFACES

> I/O

MODULE

> PLC

OUTPUTS

>

C

ONFIGURATION

DIG

. O

UTPUT

WARNING

closed:
open:

Warning
no warning

ERROR

closed:
open:

Error
no error

CAL_ACTIVE

closed:
open:

Unit is calibrated.
Unit is not calibrated.

CAL_REQUEST

closed:

closed:

open:

and no external calibration: Calibration request
(e. g. in case of temperature change of 5 °C)
and external calibration: Request "Open external calibration
leak"
no request

RUN_UP

closed:
open:

Running up
no run-up

ZERO_ACTIVE

closed:
open:

Z

ERO

switched on

Z

ERO

switched off

EMISSION_ON

closed:
open:

Emission switched on
Emission switched off

MEASURE

closed:

open:

Meas
(Z

ERO

is possible, all trigger outputs switch depending on the

leakage rate.)
Standby or Emission disabled
(Z

ERO

is not possible, all trigger outputs will return "Fail".)

STANDBY

closed:

open:

Standby
(Z

ERO

is not possible, all trigger outputs will return "Fail".)

Meas
(Z

ERO

is possible, all trigger outputs switch depending on the

leakage rate.)

SNIFF

closed:
open:

SNIFF
VAC

Function

State:

Description