Operation
252
E
Voltage REF5V
shortened,
(voltage [V])
24
120
+15V or REV5V too low,
short circuit at pre-
amplifier, IF board or MSB
defective
3xx detection system (offset pre-amplifier, pre-amplifier test, emission, cathode test)
300
W
Anode voltage too low,
(anode voltage [V])
41
132
7V <
Nom.
Value
Short circuit anode voltage,
pressure in mass
spectrometer too high, IF
board, MSB or ion source
defective
301
W
Anode voltage too high,
(anode voltage [V])
40
131
7V >
Nom.
Value
MSB defective
302
W
Suppressor voltage too
low,
(suppressor voltage [V])
39
130
297V
Short circuit suppressor, IF
board or MSB defective
303
W
Suppressor voltage too
high,
(suppressor voltage [V])
38
129
363V
MSB defective
304
W
Anode cathode voltage
too low,
(anode - cathode
voltage)
36
127
40V
Short circuit anode-
cathode, IF board or MSB
defective
305
W
Anode cathode voltage
too high,
(anode - cathode
voltage)
35
126
140V
MSB defective
310
W
Cathode 1 broken,
(actual status)
45
136
Cathode defective, line to
cathode interrupted, IF
board or MSB defective
311
W
Cathode 2 broken,
(actual status)
46
137
Cathode defective, line to
cathode interrupted, IF
board or MSB defective
312
E
Cathode broken,
(actual status)
47
138
Cathode defective, line to
cathode interrupted, IF
board or MSB defective
340
E
Emission error,
(emission current [A])
44
135
<90%
Nom.
value
>110%
Nom.
Value
Emission was stable
previously, pressure
probably too high,
message after 15s
342
W
Cathodes not
connected
47
138
Both cathodes defective
during self-testing or plug
not connected
350
W
Suppressor not
connected
39
130
Suppressor cable during
self-testing not connected
or defective
Error
number
Warning/
Error
Error message
LDS3000
Error
number
LDS1000
protocol
Error
number
ASCII
protocol*
Limit
values
Cause