Operation

252

E

Voltage REF5V
shortened,
(voltage [V])

24

120

+15V or REV5V too low,
short circuit at pre-
amplifier, IF board or MSB
defective

3xx detection system (offset pre-amplifier, pre-amplifier test, emission, cathode test)

300

W

Anode voltage too low,
(anode voltage [V])

41

132

7V <
Nom.
Value

Short circuit anode voltage,
pressure in mass
spectrometer too high, IF
board, MSB or ion source
defective

301

W

Anode voltage too high,
(anode voltage [V])

40

131

7V >
Nom.
Value

MSB defective

302

W

Suppressor voltage too
low,
(suppressor voltage [V])

39

130

297V

Short circuit suppressor, IF
board or MSB defective

303

W

Suppressor voltage too
high,
(suppressor voltage [V])

38

129

363V

MSB defective

304

W

Anode cathode voltage
too low,
(anode - cathode
voltage)

36

127

40V

Short circuit anode-
cathode, IF board or MSB
defective

305

W

Anode cathode voltage
too high,
(anode - cathode
voltage)

35

126

140V

MSB defective

310

W

Cathode 1 broken,
(actual status)

45

136

Cathode defective, line to
cathode interrupted, IF
board or MSB defective

311

W

Cathode 2 broken,
(actual status)

46

137

Cathode defective, line to
cathode interrupted, IF
board or MSB defective

312

E

Cathode broken,
(actual status)

47

138

Cathode defective, line to
cathode interrupted, IF
board or MSB defective

340

E

Emission error,
(emission current [A])

44

135

<90%
Nom.
value
>110%
Nom.
Value

Emission was stable
previously, pressure
probably too high,
message after 15s

342

W

Cathodes not
connected

47

138

Both cathodes defective
during self-testing or plug
not connected

350

W

Suppressor not
connected

39

130

Suppressor cable during
self-testing not connected
or defective

Error

number

Warning/

Error

Error message
LDS3000

Error

number

LDS1000

protocol

Error

number

ASCII

protocol*

Limit
values

Cause