Assigning the digital inputs of the I/O module
The available functions can be assigned in any way necessary to the digital inputs
PLC-IN 1 ... 10 of the I/O module.
Active signal: typically 24 V; inactive signal: typically 0 V.
The 24V output of the I/O module can be used as an active signal.
26
2
Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-9 mbar·l/s
Leak rate log.
1 V/dec.
1E1 mbar·l/s
27
1
Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-7 mbar·l/s
Leak rate linear
irrelevant 1E-6 mbar·l/s
27
2
Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-10 mbar·l/s
Leak rate log.
1 V/dec.
1E0 mbar·l/s
28
1
Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-8 mbar·l/s
Leak rate linear
irrelevant 1E-7 mbar·l/s
28
2
Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s
Leak rate log.
1 V/dec.
1E-1 mbar·l/s
29
1
Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-9 mbar·l/s
Leak rate linear
irrelevant 1E-8 mbar·l/s
29
2
Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s
Leak rate log.
1 V/dec.
1E-2 mbar·l/s
30
1
Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-10 mbar·l/s
Leak rate linear
irrelevant 1E-9 mbar·l/s
30
2
Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s
Leak rate log.
1 V/dec.
1E-3 mbar·l/s
LDS2010
setting in
menu item
22
Analog
output
LDS2010 function
LDS3000 function
Decade
scale
Upper limit
value
(10 V = ...)
Function
Flank/state:
Description
NO_FUNCTION
-
no function
DYN_CAL
inactive→ active:
active→ inactive:
Start external dynamic calibration.
Apply value for background and finish calibration.
CAL_EXTERN
inactive→ active:
active→ inactive:
Start external calibration.
Apply value for background and finish calibration.
CAL_INTERN
inactive→ active: Start internal calibration.
SNIFF/VAC
inactive→ active:
active→ inactive:
Enable sniffer mode.
Enable vacuum mode.
START
inactive→ active: Switch to Meas.
(Z
ERO
is possible, all trigger outputs switch depending on the
leakage rate.)