Operation

27

6.3

Assigning the digital inputs of the I/O module

The available functions can be assigned in any way necessary to the digital inputs
PLC-IN 1 ... 10 of the I/O module.

Active signal: typically 24 V; inactive signal: typically 0 V.
The 24V output of the I/O module can be used as an active signal.

26

2

Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-9 mbar·l/s

Leak rate log.

1 V/dec.

1E1 mbar·l/s

27

1

Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-7 mbar·l/s

Leak rate linear

irrelevant 1E-6 mbar·l/s

27

2

Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-10 mbar·l/s

Leak rate log.

1 V/dec.

1E0 mbar·l/s

28

1

Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-8 mbar·l/s

Leak rate linear

irrelevant 1E-7 mbar·l/s

28

2

Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s

Leak rate log.

1 V/dec.

1E-1 mbar·l/s

29

1

Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-9 mbar·l/s

Leak rate linear

irrelevant 1E-8 mbar·l/s

29

2

Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s

Leak rate log.

1 V/dec.

1E-2 mbar·l/s

30

1

Leak rate lin. in mbar·l/s
0 ... 10 V, 1 V = 1E-10 mbar·l/s

Leak rate linear

irrelevant 1E-9 mbar·l/s

30

2

Leak rate log. in mbar·l/s
0 ... 4 V, 1 V/decade,
0 V = 1E-11 mbar·l/s

Leak rate log.

1 V/dec.

1E-3 mbar·l/s

LDS2010
setting in
menu item
22

Analog
output

LDS2010 function

LDS3000 function

Decade
scale

Upper limit
value
(10 V = ...)

Function

Flank/state:

Description

NO_FUNCTION

-

no function

DYN_CAL

inactive→ active:
active→ inactive:

Start external dynamic calibration.
Apply value for background and finish calibration.

CAL_EXTERN

inactive→ active:
active→ inactive:

Start external calibration.
Apply value for background and finish calibration.

CAL_INTERN

inactive→ active: Start internal calibration.

SNIFF/VAC

inactive→ active:
active→ inactive:

Enable sniffer mode.
Enable vacuum mode.

START

inactive→ active: Switch to Meas.

(Z

ERO

is possible, all trigger outputs switch depending on the

leakage rate.)